DESIGN no.2
APRIL 2000

Mainly for RF reasons the prototype design 1 is 'reshaped'. The two detector boxes are mounted together in a centerframe before installation in the vacuumtank. This makes it possibile to install WF suppressors along the sides. There is one feedthrough flange at one side of the vacuumtank.

Comments:

Plus points:
    Center frame much stiffer optimally positioned short legs, no more problems of bending not any longer 'in our way' (more accessibility to critical items of VD, see below) All 'referencing' is achieved with a much simpler design alignment and reproducibility VD easier to mount and position in the tank build in complete VD at once solves accessibility problem on one side of the beam pipe no more need of tilting mechanism during VD installation Easier access to the critical WF suppressor more accessibility for connection to exit foil Possibility to install WF suppressors along detector box sides required for position 'open' !! Much space left on one side of the tank secondary vacuum pumps and, if required, a titanium evaporator with much more favorable design conditions (comes from side, instead of from front of tank) Pumping on the detector boxes is more efficient
Minus points:
    re-do FEA calculations two detector boxes are not interchangeable any longer (but individual stations can be) all feedthroughs on the same big flange all cables and repeater cards on same side present choice of feedthrough limits total number of pins to about 25000 !!

Pictures:

Total design
Si detectors
Vacuum Tank
Si detector
Center frame