The Monte Carlo data can be used to show the detector resolution
for the CC events. Figure 4.1(a) and (b) show
the detector response for
and
respectively.
The reconstructed value
is shown on the vertical, the true value on the horizontal axis.
Clearly the reconstructed value is systematically lower than
the true value. This stems from the fact that the particles have
to pass through inactive material before reaching the UCAL
and thus lose energy.
Figure 4.2(a) shows the
resolution
as function of
.
The resolution is about for
small
(
) and
less than for larger
values.
Figure 4.2(b) shows the quality of the vertex
reconstruction - only events for which a vertex is reconstructed
are shown. The vertex resolution is about
.